The TE-PUFPLUS is a high volume air sampler specifically engineered for polyurethane foam (PUF) sampling of volatile organic compounds (VOCs) and airborne pesticides in ambient air. This system meets EPA requirements for methods TO-4A, TO-9A, and TO-13A, making it compliant for regulatory ambient air quality monitoring programs.
Dual-Phase Collection Technology
The sampler uses a dual sampling module that collects airborne particulates onto filter paper media in the upper chamber while capturing vapor-phase contaminants onto polyurethane foam in the lower chamber. This configuration enables comprehensive collection of both particle-bound and gaseous organic compounds in a single sampling event.
PLUS Controller Technology
The PLUS technology flow controller uses ambient temperature, ambient pressure, and orifice differential pressure to maintain the set flow rate, providing the most accurate flow control available. Key features include:
- Built-in ambient temperature and barometric pressure sensors
- Auto-calibration function for field reliability
- Digital timer with battery backup
- Four data logs capturing average, minimum, and maximum ambient conditions
- USB export capability for data retrieval
- Total ambient and standard flow recording
Specifications
- Power: 110V, 60Hz
- Motor Type: Brush-style motor assembly
- Controller: MFC PLUS with microprocessor-based control
- Timer: Digital with programmable sampling schedules
- Flow Rate: 125-250 liters per minute (lpm) operating range
Motor and Maintenance
The brush-style motor uses carbon motor brushes to conduct current within the electric motor. While low cost and field-proven, these motors require routine motor brush changes as part of regular maintenance intervals. This configuration provides reliable performance for standard ambient monitoring applications where periodic maintenance access is available.
Applications
Designed for ambient air quality monitoring programs requiring EPA-compliant VOC and pesticide sampling. The sampling module consists of a filter and glass sampling cartridge containing PUF to concentrate PAHs and other organic compounds from air. Typical applications include:
- Regulatory ambient air monitoring for VOCs and pesticides
- EPA TO-4A, TO-9A, and TO-13A compliance sampling
- Environmental assessment studies
- Source emission characterization
- Controller Type
- MFC PLUS
- Motor Type
- Brush
- Power Configuration
- 110volt, 60hz
- Sample Type
- PUF
- Timer Type
- Digital
- Uses PLUS tech?
- Yes