Secure Checkout
TSP Hi-Vol Air Sampler - MFC, Mechanical Timer, Brush Motor, 220V 60Hz

Thumbnail Filmstrip of TSP Hi-Vol Air Sampler - MFC, Mechanical Timer, Brush Motor, 220V 60Hz Images

    Purchase TSP Hi-Vol Air Sampler - MFC, Mechanical Timer, Brush Motor, 220V 60Hz

    TSP Hi-Vol Air Sampler - MFC, Mechanical Timer, Brush Motor, 220V 60Hz SKU: TE-5170XZ

    Out of Stock

    The TE-5170XZ is a high volume air sampler configured for Total Suspended Particulate (TSP) sampling in accordance with 40 CFR Part 50, Appendix B. This EPA reference method provides measurement of total suspended particulate matter in ambient air for determining compliance with national ambient air quality standards.

    Mass Flow Control Technology

    The integrated mass flow controller (MFC) maintains consistent sampling flow rates by automatically adjusting motor speed based on real-time flow feedback from a hot wire anemometer probe. The hot wire anemometer probe is inserted into the flow stream and automatically adjusts motor speed as the filter begins to collect particulate, compensating for filter loading throughout the 24-hour sampling period. The system samples between 40-60 ft³/min (1.1-1.7 m³/min) of air with the sampling rate held constant over the sampling period.

    Key Specifications

    • Power: 220V, 60Hz configuration
    • Motor: Brush-style drive system
    • Timer: 7-day mechanical timer with dependable start/stop control
    • Particle collection: Favors particles up to 25-50 µm aerodynamic diameter depending on wind conditions
    • Sampling period: 1,440 ±60 minutes for valid sample

    System Components

    Housed in a rugged anodized aluminum shelter with gabled roof design that allows ambient air to enter from all four sides. The shelter is designed to allow particulate matter to be drawn on to the filter from all four sides. The mechanical timer provides simple, field-proven operation without requiring complex programming or digital interfaces.

    Compatible Configurations

    The TE-5170XZ represents one configuration within the TE-5170 series. Alternative configurations include:

    • TE-5170 - 110V 60Hz, brush motor, MFC
    • TE-5170BLXZ - 220V 60Hz, brushless motor, MFC
    • TE-5170VXZ - 220V 60Hz, brush motor, VFC
    • Digital timer variants available (TE-5170DXZ series)

    Applications

    Used for determining compliance with primary and secondary national ambient air quality standards for particulate matter. Ideal for ambient air monitoring networks, environmental research, and regulatory compliance monitoring. The measurement process is nondestructive, and collected samples are adequate for subsequent chemical analysis including metals analysis for comprehensive air quality assessment.

    Maintenance Notes

    Brush motor systems require periodic brush replacement as part of routine maintenance. The elapsed time indicator tracks motor life and brush changes to verify sampling time. Standard maintenance includes filter replacement, shelter cleaning, and flow calibration verification according to Tisch Environmental's recommended service intervals.

    Controller Type
    MFC
    Motor Type
    Brush
    Power Configuration
    220volt, 60hz
    Sample Type
    TSP
    Timer Type
    Mechanical