Chat with us, powered by LiveChat

Secure Checkout
GD/X 25mm Syringe Filter, Glass Microfiber GF/A Filteration Medium, 1.6um (150 pcs)

Thumbnail Filmstrip of GD/X 25mm Syringe Filter, Glass Microfiber GF/A Filteration Medium, 1.6um (150 pcs) Images

    Purchase GD/X 25mm Syringe Filter, Glass Microfiber GF/A Filteration Medium, 1.6um (150 pcs)

    GD/X 25mm Syringe Filter, Glass Microfiber GF/A Filteration Medium, 1.6um (150 pcs)
    Whatman SKU: 6882-2516

    $646.99
    Ships in 4 business days
     

    Whatman GD/X Glass Fiber (GF) Syringe Filters, 1.6 µm, 25 mm, 150/Pk - 6882-2516

    Whatman ties GD/X prefiltration performance to one closely held fiber-grade standard applied at every production run, keeping results repeatable from one manufacturing lot to the next. Syringe filters span a wide range of membrane chemistries and pore sizes, so the format can be matched to the sample matrix and solvent system in use. Inorganic glass fiber offers broad chemical compatibility for demanding samples while wetting rapidly with aqueous samples and captures particulates throughout its depth. The polypropylene housing holds its seal true run after run and routes flow to minimize backpressure in service. Where analytical results depend on clean samples, the preparation step stays consistent from sample to sample through a pre-tested syringe filter.

    Membrane Characteristics

    • High wet strength of glass fiber maintains integrity when saturated with aqueous or organic media
    • Reduces bioburden and particulate load ahead of sterilizing-grade filtration
    • Inert glass surface preserves sample pH and ionic strength throughout filtration
    • Surface silanol groups attract water molecules for immediate aqueous wetting
    • Inorganic matrix withstands repeated aqueous exposures without swelling or structural change

    Housing Characteristics

    • Minimal ionic background from the polypropylene housing supports ion chromatography and trace-metal ICP-MS sample preparation
    • Standardized luer connection geometry provides secure engagement with standard syringe tips
    • Minimal extractable release from the housing supports sensitive analytical work downstream
    • Housing tolerates elevated-temperature aqueous and solvent sample filtration up to about 80°C without softening

    A particulate-laden aqueous sample makes the rest of your preparation unpredictable. Consistent aqueous clarification clears particulates without slowing your bench routine the same way every time you run it. Request available pack sizes and lead times across this filtration line's membrane options from the Tisch Scientific team.

    Sterilization Guidance

    The housing supports customer autoclave sterilization at 121°C for 30 minutes.

    For more information about chemical compatibility, please review our Chemical Compatibility Chart.

    Connection Type
    Luer-Lok/ Luer Slip
    Construction
    Glass Microfiber (Binder Free)
    Diameter
    25mm
    Filtration Area
    4.91cm²
    Flow Rate
    Fast
    Material
    Glass Micro Fiber
    Maximum Operating Temperature (°C)
    550°C
    Membrane Thickness (µm)
    260µm
    Pack Count
    150
    Pore Size (µm)
    1.6
    Prefilter (Yes/No)
    No
    Sterility
    Nonsterile
    Sterilization
    Autoclavable 500°C
    Volume Throughput (ml)
    Up to 100ml
    Wettability
    Hydrophilic
    • Groundwater Testing - Combines an integrated glass prefilter with a glass final stage in one device, clarifying turbid groundwater samples.
    • Urine Sample Prep - Glass fiber removes urinary sediment and mucus from urine samples ahead of HPLC and immunoassay clinical testing.
    • Coatings Quality Testing - Glass fiber's chemical inertness handles viscous, solvent-laden coating formulations without swelling or degrading.
    • Analysis Prep - Ensures reliable clarification in multi-stage filtration stacks, effectively removing gross debris and precipitates safely.
    • Protein Aggregate Removal - Glass fiber removes large protein aggregates and precipitates from crude lysate ahead of activity assays and separation.
    • Air Sampling Extract Prep - Glass fiber removes insoluble particulates from air filter extracts ahead of metals and organic compound analysis.