Tisch Scientific Glass Fiber (GF) Syringe Filters, 1 µm, 25 mm, 100/Pk - GD17171
Tisch Scientific works with established filtration manufacturers to ensure reliable quality and performance. Syringe filters give the analyst a single-use device for clarifying samples right at the bench. Naturally hydrophilic glass fiber wets rapidly with aqueous samples while tolerating repeated high-temperature conditioning cycles and resists chemical attack across diverse solvents. The PP housing keeps the fluid path low in extractable contamination while preventing sample bypass under rated operating pressure and maintains a low-dead-volume internal flow path. When sterile or particle-free preparation matters at the bench, the syringe filter delivers it with minimal added handling.
Technical Specifications
- Inorganic construction suits demanding sample-preparation chemistries
- Nominal depth retention with high dirt-loading capacity
- Naturally hydrophilic surface wets rapidly with aqueous samples
- Biologically inert borosilicate fiber with broad chemical compatibility
- Withstands high-temperature pre-conditioning up to 500°C
Housing Specifications
- Resists aqueous acids, bases, and alcohols in routine sample prep
- Tough polypropylene body tolerates repeated handling at the bench
- Resists stress cracking from aqueous salt and buffer solutions
- Plasticizer-free polyolefin construction keeps the fluid path clean
Filtering particulates upstream protects the column investment your methods depend on. Upstream buffer clarification keeps avoidable particulates out of your flow path while keeping the filtrate analytically clean. Reach out to the Tisch Scientific team for application support, volume pricing, and technical assistance with your filtration product selection.
For more information about chemical compatibility, please review our Chemical Compatibility Chart.
- Connection Type
- Luer-Lok/Luer Slip
- Construction
- Polypropylene (PP)
- Diameter
- 25mm
- Filtration Area
- 4.08cm²
- Flow Rate
- 350ml
- Holdup Volume
- 400µl
- Material
- Glass Fiber (GF)
- Maximum Operating Pressure
- 87psi
- Maximum Operating Temperature (°C)
- 60°C
- Membrane Thickness (µm)
- 250-290µm
- Pack Count
- 100
- Pore Size (µm)
- 1.0
- Prefilter (Yes/No)
- Yes
- Prefilter Material
- GF+PP+GF
- Sterility
- Nonsterile
- Sterilization
- Autoclave 121°C for 30min/ gamma 25kGy
- Volume Throughput (ml)
- 50ml
- Weight
- 0.70
- Wettability
- Hydrophilic
- Environmental Water Testing – Remove sediment from turbid field samples before trace metal analysis while protecting analytical instrumentation
- Fermentation Broth Processing – Separate cells and debris from bioreactor samples before HPLC or spectroscopic measurements in bioprocess monitoring
- Industrial Fluid Analysis – Clarify hydraulic oils and coolants containing suspended contaminants before particulate analysis in quality control
- Agricultural Sample Prep – Filter plant tissue homogenates and soil extracts before pesticide residue quantification in food safety testing
- Paint and Coating Testing – Remove pigment aggregates from viscous formulations before rheology measurements or quality assurance analysis
- Tissue Homogenate Clarification – Process biological extracts containing cellular debris before protein assays or enzyme measurements in research