Chat with us, powered by LiveChat

Secure Checkout
Acro Disc Syringe Filters w/ wwPTFE Membrane - 0.45um, 13mm, Minispike Outlet, 300/Pk

Thumbnail Filmstrip of Acro Disc Syringe Filters w/ wwPTFE Membrane - 0.45um, 13mm, Minispike Outlet, 300/Pk Images

    Purchase Acro Disc Syringe Filters w/ wwPTFE Membrane - 0.45um, 13mm, Minispike Outlet, 300/Pk

    Acro Disc Syringe Filters w/ wwPTFE Membrane - 0.45um, 13mm, Minispike Outlet, 300/Pk
    Pall SKU: 2402

    $918.95
    Ships in 7 weeks
     

    Pall Acrodisc PTFE Syringe Filters, 0.45 µm, 13 mm, 300/Pk - 2402

    Syringe filter production at Pall answers to the same quality infrastructure standing behind its process-scale filtration systems. Syringe filters span a wide range of membrane chemistries and pore sizes, so the format can be matched to the sample matrix and solvent system in use. The PTFE membrane clarifies solvent streams at the rated pore size and suits solvent filtration where other membranes fail. The polypropylene housing resists aqueous acids, bases, and alcohols in routine use and preserves sample chemistry across the flow path. Where analytical results depend on clean samples, the preparation step stays consistent from sample to sample through a pre-tested syringe filter.

    Membrane Characteristics

    • Solvent tolerance includes full-strength tetrahydrofuran and chlorinated solvents such as dichloromethane
    • High membrane porosity, 70 to 80 percent void volume, delivers high flow at low pressure
    • High porosity minimizes hold-up volume for small sample volumes
    • Non-fibrous, fully synthetic structure sheds no fibers into filtrate
    • Inert composition offers no nutrient source for microbial growth

    Housing Characteristics

    • Thermal-bonded construction keeps adhesives out of the fluid path
    • Minispike inlet penetrates rubber vial septa to enable closed, contamination-free transfer from sealed source containers
    • Withstands standard 121°C steam autoclave cycles without softening or structural distortion
    • Resists stress cracking from aqueous salt and buffer solutions

    A particulate-laden aqueous sample makes the rest of your preparation unpredictable. General-purpose aqueous filtration supports the sample types you handle every day from reagent and buffer preparation through routine sample clarification. Request details on payment terms or standing-order arrangements for your routine filtration purchases.

    Sterilization Guidance

    The housing supports customer autoclave sterilization at 121°C for 30 minutes.

    For more information about chemical compatibility, please review our Chemical Compatibility Chart.

    Bubble Point (psi)
    9psi (IPA)
    Connection Type
    Luer-Lock/ Minispike
    Construction
    Polypropylene (PP)
    Diameter
    13mm
    Filtration Area
    1.0cm²
    Flow Rate
    224ml
    Holdup Volume
    < 14 µL
    Material
    Polytetrafluoroethylene (PTFE)
    Maximum Operating Pressure
    60psi
    Pack Count
    300
    Pore Size (µm)
    0.45
    Sterility
    Nonsterile
    Sterilization
    Autoclavable at 121°C for 15-20 min
    Wettability
    Hydrophobic
    • Agrochemical Formulation QC - Clarifies technical pesticide concentrates in organic solvent ahead of HPLC potency and impurity analysis.
    • Polymer GPC Prep - Clarifies polymer solutions in THF ahead of gel permeation chromatography, protecting columns from particulates.
    • Air Monitoring - Ensures precise capture of airborne particulates for environmental analysis in demanding atmospheric monitoring conditions.
    • Chemical Synthesis - Resists aggressive organic solvents and corrosive acids for compatibility with harsh synthetic reaction mixtures.
    • Cryogenic Fluids - Delivers consistent performance across extreme temperature ranges, filtering cryogenic liquids with dependable, reliable results.
    • Sample Protection - Low-binding properties ensure maximum recovery of low-concentration analytes, improving accuracy in quantitative assay results.