Pall Acrodisc PTFE Syringe Filters, 0.45 µm, 25 mm, 1000/Pk - AP-4302
Pall's proprietary membrane manufacturing capabilities put specialized media into the Acrodisc housing geometry. A syringe filter's small hold-up volume limits sample loss, a meaningful difference when working with a limited or precious sample. PTFE delivers absolute particle cutoff at the rated pore size while repelling water and passing gases and solvents freely and withstands aggressive organics that degrade other media. The PP housing stays inert across common laboratory solvents and buffers in routine use while keeping sample purity intact through the filtration pass and sustains efficient flow on viscous samples. Where small-volume samples must reach the instrument clean, low hold-up volume keeps sample loss minimal while the instrument stays protected from particulate damage.
Prefiltration
- Two-stage construction pairs a borosilicate glass fiber prefilter with the downstream membrane
- Depth-loading glass fiber matrix captures particulate throughout its full thickness
- Glass fiber prefilter stage sustains flow through particle-laden, hard-to-filter samples
- Added glass fiber stage extends usable throughput by delaying membrane clogging
Membrane Characteristics
- Fully fluorinated carbon backbone resists ultraviolet and ozone degradation
- Low non-specific protein binding preserves recovery of biomolecules and therapeutic proteins during filtration
- Low extractables keep the chromatographic background clean
- UL 94 V-0 rating confirms the highest self-extinguishing classification
- Surface energy near 18.5 dynes per centimeter repels oils and nonpolar liquids
Housing Characteristics
- Thermal-bonded construction keeps adhesives out of the fluid path
- Inert polyolefin wall contributes minimal sample adsorption to the filtrate
- Housing tolerates elevated-temperature aqueous and solvent sample filtration up to about 80°C without softening
- Polypropylene density near 0.90 g/cm3 keeps housing weight low relative to denser engineering plastics
Particulates left in your organic solvents and reagents carry downstream into the step that depends on a clean feed. Solvent-stream filtration handles the solvents and reagents your methods rely on from initial prep through final use. Start a live chat with the Tisch Scientific team about packaging, unit count, and ship-date for your filtration order.
Sterilization Guidance
Standard autoclave cycles at 121°C for 30 minutes prepare the device for sterile workflows.
For more information about chemical compatibility, please review our Chemical Compatibility Chart.
- Bubble Point (psi)
- 9psi (IPA)
- Burst Pressure (psi)
- 60psi
- Connection Type
- Luer-Lock/ Luer Slip
- Construction
- Polypropylene (PP)
- Diameter
- 25mm
- Filtration Area
- 3.9cm²
- Flow Rate
- 130ml
- Holdup Volume
- < 100 µL
- Material
- Polytetrafluoroethylene (PTFE)
- Maximum Operating Pressure
- 75psi
- Maximum Operating Temperature (°C)
- 100°C
- Pack Count
- 1000
- Pore Size (µm)
- 0.45
- Sterility
- Nonsterile
- Sterilization
- Autoclavable at 121°C for 15-20 min
- Wettability
- Hydrophobic
- Photoresist Filtration - Clarifies photoresist solutions in organic casting solvent ahead of spin-coating, reducing particulate defect risk.
- Spirits Congener Analysis - Clarifies distilled spirits samples ahead of GC analysis, protecting columns from residue during congener profiling.
- Ergonomic Filtration - Larger surface area reduces the hand force needed for manual filtration, helping minimize operator fatigue and strain.
- Battery Electrolyte Prep - Clarifies lithium-ion battery electrolyte in carbonate solvents ahead of GC-MS purity and impurity analysis.
- Sample Protection - Low-binding properties ensure maximum recovery of low-concentration analytes, improving accuracy in quantitative assay results.
- Trace Metals Analysis - PTFE's low metal-extractable profile clarifies acid-digested samples for clean, low-background ICP-MS and ICP-OES results.