Tisch Scientific reduces per-unit costs for laboratories that place blanket purchase agreements or large-quantity orders. Each hydrophobic vent-rated filter carries manufacturer bacterial retention testing built into the design, letting a vent line stay in continuous service without repeating that testing on site between uses. The PVDF membrane tolerates common chromatography solvents without degradation while blocking aqueous ingress in venting applications and achieves sterile-grade retention at sub-micron pore sizes. The polypropylene housing presses snugly against the membrane during the pass while standing firm against aqueous acids, bases, and alcohols during regular handling and sustains efficient flow on viscous samples. A vent or gas line stays free of airborne bacteria once a hydrophobic, sterilizing-grade membrane is in the path.
Prefiltration
- Depth media that capture particulate through their full thickness
- Layered glass fiber and polypropylene depth media
- Nominal depth rating sized for coarse particulate capture
- High dirt-holding capacity for the heaviest particulate loads
Membrane Characteristics
- Vapor-phase hydrogen peroxide decontamination cycles leave the membrane intact
- Low ionic extractables preserve baseline stability in conductivity-based ion chromatography detection
- Absolute pore rating ensures reproducible results between test runs
- Bacterial challenge testing with Brevundimonas diminuta at 10 million CFU per cm2 confirms sterilizing grade
- Organic solvent resistance beats PES or nylon at comparable pore size
Housing Characteristics
- Polypropylene density near 0.90 g/cm3 keeps housing weight low relative to denser engineering plastics
- Standardized luer connection geometry provides secure engagement with standard syringe tips
- Stable across the full aqueous pH range in routine sample work
- The seal formed against the membrane blocks bypass at rated operating pressure
An aggressive solvent can leach or break down a vulnerable filter and carry debris into your prep. Solvent-stream filtration handles the solvents and reagents your methods rely on so your filter holds its integrity through the run. Request a quality certificate, lot documentation, or a blanket-order quote for your filtration order at any time.
Sterilization Guidance
Customer autoclave compatibility at 121°C accommodates in-lab sterilization timing.
For more information about chemical compatibility, please review our Chemical Compatibility Chart.
- Bubble Point (psi)
- 14.5psi
- Connection Type
- Luer-Lok/Luer Slip
- Diameter
- 25mm
- Filtration Area
- 4.08cm²
- Flow Rate
- 20ml
- Holdup Volume
- 100μL
- Material
- Polyvinylidene Fluoride (PVDF)
- Maximum Operating Pressure
- 87psi
- Maximum Operating Temperature (°C)
- 60°C
- Membrane Thickness (µm)
- 100-130μm
- Pack Count
- 100
- Pore Size (µm)
- 0.22
- Prefilter (Yes/No)
- Yes
- Prefilter Material
- GF+GF+PP
- Sterility
- Nonsterile
- Sterilization
- Autoclave at 121°C for 30min/ gamma 25kGy
- Volume Throughput (ml)
- 50ml
- Weight
- 0.70
- Wettability
- Hydrophobic
- Bioreactor Vent Protection - Repels water while venting bioreactors and fermenters, keeping process air and gases moving freely through the line.
- Culture Vessel Filling - Equalizes pressure in sealed culture vessels by venting displaced air, keeping fill and drain lines fully closed.
- Dissolution Testing - Removes undissolved particles from timed USP 711 dissolution aliquots, preventing skewed potency and release results.
- Sterile Tank Venting - Blocks aqueous ingress on storage tanks and carboys, letting sterile air and gas pass through the vent line.
- QuEChERS Extract Prep - Removes co-extracted waxes and pigments from QuEChERS extracts, protecting GC-MS and LC-MS systems during pesticide testing.
- Vial and Tubing Venting - Blocks liquid while venting sealed vials and tubing lines, keeping pressure equalized during fluid transfer steps.