    Tisch Scientific's representatives greet every caller with patience, whatever brings them to the phone. Each hydrophobic vent-rated filter carries manufacturer bacterial retention testing built into the design, letting a vent line stay in continuous service without repeating that testing on site between uses. The hydrophobic PTFE membrane suits gas, vent, and solvent-stream filtration and delivers sterilizing-grade gas and vent filtration. The PP housing stays inert across common laboratory solvents and buffers in routine use while maintaining a low-dead-volume internal flow path and presses snugly against the membrane during the pass. Downstream equipment and work areas stay protected from airborne bacterial contamination once the vent path runs through the membrane.

    Membrane Characteristics

    • Non-wetting pore surface blocks liquid water while passing water vapor freely
    • Chemical resistance spans the full pH 1 to 14 range
    • Membrane integrity holds across repeated autoclave sterilization cycles at 121°C to 134°C
    • Solvent tolerance includes full-strength tetrahydrofuran and chlorinated solvents such as dichloromethane
    • UL 94 V-0 rating confirms the highest self-extinguishing classification

    Housing Characteristics

    • The seal formed against the membrane blocks bypass at rated operating pressure
    • Medical-grade polypropylene with broad chemical compatibility
    • Burst pressure rating provides a wide safety margin above the rated maximum operating pressure
    • Minimal ionic background from the polypropylene housing supports ion chromatography and trace-metal ICP-MS sample preparation

    An aggressive solvent can leach or break down a vulnerable filter and carry debris into your prep. Solvent-stream filtration handles the solvents and reagents your methods rely on across the organic solvents and harsh reagents your methods use. Request a quote on bulk pricing for larger filtration orders from the Tisch Scientific team.

    Sterilization Guidance

    The housing supports customer autoclave sterilization at 121°C for 30 minutes.

    For more information about chemical compatibility, please review our Chemical Compatibility Chart.

    Bubble Point (psi)
    26.11psi
    Burst Pressure (psi)
    87psi.
    Connection Type
    Luer-Lok/Luer Slip
    Construction
    Polypropylene (PP)
    Diameter
    33mm
    Filtration Area
    4.6cm²
    Flow Rate
    12ml
    Holdup Volume
    100μL
    Incompatibility
    Aqueous samples without rewetting
    Material
    Polytetrafluoroethylene (PTFE)
    Maximum Operating Pressure
    87psi
    Maximum Operating Temperature (°C)
    60°C
    Membrane Thickness (µm)
    80-140μm
    Pack Count
    100
    Pore Size (µm)
    0.1
    Sterility
    NonSterile
    Sterilization
    Autoclave 121°C for 30min/ Gamma 25kGy
    Volume Throughput (ml)
    120ml
    Weight
    0.96
    Wettability
    Hydrophobic
    • Chemical Synthesis - Resists aggressive organic solvents and corrosive acids for compatibility with harsh synthetic reaction mixtures.
    • Air Monitoring - Ensures precise capture of airborne particulates for environmental analysis in demanding atmospheric monitoring conditions.
    • Dissolution Testing - Stops the dissolution reaction and clarifies samples, with low API binding for accurate potency quantitation.
    • Extractables Testing - Clarifies packaging material organic solvent extracts ahead of LC-MS analysis, keeping background low for leachable detection.
    • Dye and Pigment Analysis - Clarifies dye solutions in DMF, NMP, and other aggressive organics ahead of spectrophotometric color strength testing.
    • Agrochemical Formulation QC - Clarifies technical pesticide concentrates in organic solvent ahead of HPLC potency and impurity analysis.